The Doppler effect is used to measure the speed of blood flow in arteries.
Which medical diagnostic method uses this technique?
barium scan
CAT scan
PET scan
ultrasound scan
Did this page help you?
The Doppler effect is used to measure the speed of blood flow in arteries.
Which medical diagnostic method uses this technique?
barium scan
CAT scan
PET scan
ultrasound scan
Did this page help you?
The table shows some data on two tissues in a patient.
Tissue | Density | Acoustic impedance |
X | ρ | 1.5 Z |
Y | 1.3 ρ | Z |
Ultrasound in tissue X is incident at the boundary between the tissues X and Y.
What is the percentage of the ultrasound intensity reflected at the boundary?
1.7 %
4.0 %
13 %
20 %
Did this page help you?
A beam of ultrasound is incident normally at a boundary between two tissues F and G.
The table below shows some data on the two tissues.
Tissue F | Tissue G | |
Density of tissue | ρ | 1.2ρ |
Speed of ultrasound in tissue | c | 1.5c |
What percentage of the intensity of the ultrasound is reflected at the boundary?
0.83%
8.2%
9.1%
29%
Did this page help you?